An easy to scale up modular PET scanner was developed for imaging small animals. Energy resolution, spatial resolution and count rate performance were determined as system parameters. The configuration provided an average energy resolution of 19.6 % and the image resolution ranges was 1.5 to 2.3 mm in radial direction. The sensitivity of the miniPET was 1.08 cps/kBq as determined using a point source. In addition, results of rat brain scan performed with FDG are given to characterize imaging capability of the system. The displayed data document that the miniPET scanner supports good quality brain imaging of small animals.

Original languageEnglish
Title of host publication2005 IEEE Nuclear Science Symposium Conference Record -Nuclear Science Symposium and Medical Imaging Conference
Pages1645-1648
Number of pages4
DOIs
Publication statusPublished - dec. 1 2005
EventNuclear Science Symposium Conference Record, 2005 IEEE - , Puerto Rico
Duration: okt. 23 2005okt. 29 2005

Publication series

NameIEEE Nuclear Science Symposium Conference Record
Volume3
ISSN (Print)1095-7863

Other

OtherNuclear Science Symposium Conference Record, 2005 IEEE
CountryPuerto Rico
Period10/23/0510/29/05

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Fingerprint Dive into the research topics of 'Performance Characteristics of a miniPET Scanner Dedicated to Small Animal Imaging'. Together they form a unique fingerprint.

  • Cite this

    Kis, S. A., Valastyán, I., Hegyesi, G., Imrek, J., Kalinka, G., Molnár, J., Novák, D., Végh, J., Balkay, L., Emri, M., Molnár, G., Bagaméry, I., Bükki, T., Rózsa, S., Szabó, Z., Kerek, A., & Trón, L. (2005). Performance Characteristics of a miniPET Scanner Dedicated to Small Animal Imaging. In 2005 IEEE Nuclear Science Symposium Conference Record -Nuclear Science Symposium and Medical Imaging Conference (pp. 1645-1648). [1596635] (IEEE Nuclear Science Symposium Conference Record; Vol. 3). https://doi.org/10.1109/NSSMIC.2005.1596635