Fast and sensitive determination of Polysorbate 80 in solutions containing proteins

Szabolcs Fekete, Katalin Ganzler, Jenodouble acute Fekete

Research output: Article

31 Citations (Scopus)

Abstract

A fast and sensitive method has been developed for the specific determination of Polysorbate 80 (Tween 80) in liquid formulations in the presence of proteins and excipients. The quantitative determination is based on a fast liquid chromatographic (HPLC) separation and charged aerosol detection (CAD). The method was validated using a Poroshell 300SB-C18 column packed with 5μm shell particles (75mm× 2.1mm) and acetonitrile-methanol-water-trifluoroacetic acid mobile phase at a flow rate of 0.65. ml/min. The rapid LC-CAD method is suitable for quantifying Polysorbate 80 in the range of 10-60μg/ml in protein solutions within good manufacturing practices (GMPs) of the pharmaceutical industry.

Original languageEnglish
Pages (from-to)672-679
Number of pages8
JournalJournal of Pharmaceutical and Biomedical Analysis
Volume52
Issue number5
DOIs
Publication statusPublished - szept. 1 2010

ASJC Scopus subject areas

  • Analytical Chemistry
  • Pharmaceutical Science
  • Drug Discovery
  • Spectroscopy
  • Clinical Biochemistry

Fingerprint Dive into the research topics of 'Fast and sensitive determination of Polysorbate 80 in solutions containing proteins'. Together they form a unique fingerprint.

  • Cite this