Stabilisierung der Herzfrequenz zur Behandlung rezidivierender VT

Translated title of the contribution: Ventricular rate stabilization for treatment of recurrent VT

Mate Vamos, Peter Bogyi, G. Duray, Noemi Nyolczas, Stefan H. Hohnloser

Research output: Contribution to journalReview article

Abstract

A patient with ischaemic cardiomyopathy received a secondary prevention VVI implantable cardioverter–defibrillator (ICD) after an episode of sustained ventricular tachycardia (VT). Because of recurrent VTs transmitted via CareLink, medical therapy was optimized and VT ablation was performed. Subsequently, a fast VT with a typical short–long–short initiation developed. In addition, there was an increasing need for ventricular pacing due to sinus bradycardia. This new type of VT could be successfully dealt with by upgrading to DDD ICD and activating the Ventricular Rate Stabilization algorithm.

Original languageGerman
Pages (from-to)239-242
Number of pages4
JournalHerzschrittmachertherapie und Elektrophysiologie
Volume28
Issue number2
DOIs
Publication statusPublished - Jun 1 2017

Fingerprint

Ventricular Tachycardia
Dichlorodiphenyldichloroethane
Bradycardia
Cardiomyopathies
Therapeutics

Keywords

  • CareLink
  • Heart failure
  • Rate control
  • Short–long–short sequence
  • Ventricular rate stabilization

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Physiology (medical)

Cite this

Stabilisierung der Herzfrequenz zur Behandlung rezidivierender VT. / Vamos, Mate; Bogyi, Peter; Duray, G.; Nyolczas, Noemi; Hohnloser, Stefan H.

In: Herzschrittmachertherapie und Elektrophysiologie, Vol. 28, No. 2, 01.06.2017, p. 239-242.

Research output: Contribution to journalReview article

Vamos, Mate ; Bogyi, Peter ; Duray, G. ; Nyolczas, Noemi ; Hohnloser, Stefan H. / Stabilisierung der Herzfrequenz zur Behandlung rezidivierender VT. In: Herzschrittmachertherapie und Elektrophysiologie. 2017 ; Vol. 28, No. 2. pp. 239-242.
@article{e2693af33dfe46b28b4b110c25efdeba,
title = "Stabilisierung der Herzfrequenz zur Behandlung rezidivierender VT",
abstract = "A patient with ischaemic cardiomyopathy received a secondary prevention VVI implantable cardioverter–defibrillator (ICD) after an episode of sustained ventricular tachycardia (VT). Because of recurrent VTs transmitted via CareLink, medical therapy was optimized and VT ablation was performed. Subsequently, a fast VT with a typical short–long–short initiation developed. In addition, there was an increasing need for ventricular pacing due to sinus bradycardia. This new type of VT could be successfully dealt with by upgrading to DDD ICD and activating the Ventricular Rate Stabilization algorithm.",
keywords = "CareLink, Heart failure, Rate control, Short–long–short sequence, Ventricular rate stabilization",
author = "Mate Vamos and Peter Bogyi and G. Duray and Noemi Nyolczas and Hohnloser, {Stefan H.}",
year = "2017",
month = "6",
day = "1",
doi = "10.1007/s00399-017-0513-3",
language = "German",
volume = "28",
pages = "239--242",
journal = "Herzschrittmachertherapie und Elektrophysiologie",
issn = "0938-7412",
publisher = "D. Steinkopff-Verlag",
number = "2",

}

TY - JOUR

T1 - Stabilisierung der Herzfrequenz zur Behandlung rezidivierender VT

AU - Vamos, Mate

AU - Bogyi, Peter

AU - Duray, G.

AU - Nyolczas, Noemi

AU - Hohnloser, Stefan H.

PY - 2017/6/1

Y1 - 2017/6/1

N2 - A patient with ischaemic cardiomyopathy received a secondary prevention VVI implantable cardioverter–defibrillator (ICD) after an episode of sustained ventricular tachycardia (VT). Because of recurrent VTs transmitted via CareLink, medical therapy was optimized and VT ablation was performed. Subsequently, a fast VT with a typical short–long–short initiation developed. In addition, there was an increasing need for ventricular pacing due to sinus bradycardia. This new type of VT could be successfully dealt with by upgrading to DDD ICD and activating the Ventricular Rate Stabilization algorithm.

AB - A patient with ischaemic cardiomyopathy received a secondary prevention VVI implantable cardioverter–defibrillator (ICD) after an episode of sustained ventricular tachycardia (VT). Because of recurrent VTs transmitted via CareLink, medical therapy was optimized and VT ablation was performed. Subsequently, a fast VT with a typical short–long–short initiation developed. In addition, there was an increasing need for ventricular pacing due to sinus bradycardia. This new type of VT could be successfully dealt with by upgrading to DDD ICD and activating the Ventricular Rate Stabilization algorithm.

KW - CareLink

KW - Heart failure

KW - Rate control

KW - Short–long–short sequence

KW - Ventricular rate stabilization

UR - http://www.scopus.com/inward/record.url?scp=85020090601&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85020090601&partnerID=8YFLogxK

U2 - 10.1007/s00399-017-0513-3

DO - 10.1007/s00399-017-0513-3

M3 - Review article

C2 - 28567490

AN - SCOPUS:85020090601

VL - 28

SP - 239

EP - 242

JO - Herzschrittmachertherapie und Elektrophysiologie

JF - Herzschrittmachertherapie und Elektrophysiologie

SN - 0938-7412

IS - 2

ER -