The results of mass testing of the plastic streamer tubes for the DELPHI hadron calorimeter

G. D. Alekseev, Yu Ye Bonushkin, S. Czellar, B. Fialovszki, B. A. Khomenko, N. N. Khovansky, A. V. Korytov, P. A. Kulinich, M. Lokajiček, G. D. Mitselmakher, J. Patočka, V. D. Ryabtsov, J. Řidký, Yu V. Sedykh, V. V. Tokmenin, R. Toledo, E. Tarjan

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The procedure and results of mass testing of plastic streamer tubes for the DELPHI hadron calorimeter are described in detail. The mass testing includes three stages: high voltage test and conditioning, X-ray test on radiation load capability and wire tension control in ready detectors. The productivity of the test workshop was 240 detectors per day at maximum. After testing of about 25 000 streamer tubes, 8% have been rejected on the ground of severe rejection criteria which exceed significantly the working conditions in the DELPHI experiment.

Original languageEnglish
Pages (from-to)344-353
Number of pages10
JournalNuclear Inst. and Methods in Physics Research, A
Volume305
Issue number2
DOIs
Publication statusPublished - Jul 20 1991

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Dive into the research topics of 'The results of mass testing of the plastic streamer tubes for the DELPHI hadron calorimeter'. Together they form a unique fingerprint.

  • Cite this

    Alekseev, G. D., Bonushkin, Y. Y., Czellar, S., Fialovszki, B., Khomenko, B. A., Khovansky, N. N., Korytov, A. V., Kulinich, P. A., Lokajiček, M., Mitselmakher, G. D., Patočka, J., Ryabtsov, V. D., Řidký, J., Sedykh, Y. V., Tokmenin, V. V., Toledo, R., & Tarjan, E. (1991). The results of mass testing of the plastic streamer tubes for the DELPHI hadron calorimeter. Nuclear Inst. and Methods in Physics Research, A, 305(2), 344-353. https://doi.org/10.1016/0168-9002(91)90550-A