Solid sampling determination of lithium and sodium additives in microsamples of yttrium oxyorthosilicate by high-resolution continuum source graphite furnace atomic absorption spectrometry

Nikoletta Laczai, L. Kovács, A. Péter, László Bencs

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (SS-HR-CS-GFAAS) methods were developed and studied for the fast and sensitive quantitation of Li and Na additives in microsamples of cerium-doped yttrium oxyorthosilicate (Y2SiO5:Ce) scintillator materials. The methods were optimized for solid samples by studying a set of GFAAS conditions (i.e., the sample mass, sensitivity of the analytical lines, and graphite furnace heating programs). Powdered samples in the mass range of 0.099-0.422 mg were dispensed onto graphite sample insertion boats, weighed and analyzed. Pyrolysis and atomization temperatures were optimized by the use of single-element standard solutions of Li and Na (acidified with 0.144 mol/L HNO3) at the Li I 610.353 nm and Na I 285.3013 nm analytical lines. For calibration purposes, the method of standard addition with Li and Na solutions was applied. The correlation coefficients (R values) of the calibration graphs were not worse than 0.9678. The limit of detection for oxyorthosilicate samples was 20 μg/g and 80 μg/g for Li and Na, respectively. The alkaline content of the solid samples were found to be in the range of 0.89 and 8.4 mg/g, respectively. The accuracy of the results was verified by means of analyzing certified reference samples, using methods of standard (solution) addition calibration.

Original languageEnglish
Pages (from-to)8-15
Number of pages8
JournalSpectrochimica Acta - Part B Atomic Spectroscopy
Volume117
DOIs
Publication statusPublished - Mar 1 2016

Fingerprint

Yttrium
Atomic absorption spectrometry
Graphite
yttrium
Lithium
furnaces
Furnaces
graphite
lithium
Sodium
sampling
sodium
Calibration
Sampling
continuums
high resolution
spectroscopy
Heating furnaces
Cerium
Boats

Keywords

  • Advanced materials
  • Alkaline elements
  • Direct solid sampling analysis
  • HR-CS-AAS
  • Refractory matrix components

ASJC Scopus subject areas

  • Instrumentation
  • Atomic and Molecular Physics, and Optics
  • Analytical Chemistry
  • Spectroscopy

Cite this

@article{a9a8ffe9d64440ccaae2731f3546cde1,
title = "Solid sampling determination of lithium and sodium additives in microsamples of yttrium oxyorthosilicate by high-resolution continuum source graphite furnace atomic absorption spectrometry",
abstract = "Solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (SS-HR-CS-GFAAS) methods were developed and studied for the fast and sensitive quantitation of Li and Na additives in microsamples of cerium-doped yttrium oxyorthosilicate (Y2SiO5:Ce) scintillator materials. The methods were optimized for solid samples by studying a set of GFAAS conditions (i.e., the sample mass, sensitivity of the analytical lines, and graphite furnace heating programs). Powdered samples in the mass range of 0.099-0.422 mg were dispensed onto graphite sample insertion boats, weighed and analyzed. Pyrolysis and atomization temperatures were optimized by the use of single-element standard solutions of Li and Na (acidified with 0.144 mol/L HNO3) at the Li I 610.353 nm and Na I 285.3013 nm analytical lines. For calibration purposes, the method of standard addition with Li and Na solutions was applied. The correlation coefficients (R values) of the calibration graphs were not worse than 0.9678. The limit of detection for oxyorthosilicate samples was 20 μg/g and 80 μg/g for Li and Na, respectively. The alkaline content of the solid samples were found to be in the range of 0.89 and 8.4 mg/g, respectively. The accuracy of the results was verified by means of analyzing certified reference samples, using methods of standard (solution) addition calibration.",
keywords = "Advanced materials, Alkaline elements, Direct solid sampling analysis, HR-CS-AAS, Refractory matrix components",
author = "Nikoletta Laczai and L. Kov{\'a}cs and A. P{\'e}ter and L{\'a}szl{\'o} Bencs",
year = "2016",
month = "3",
day = "1",
doi = "10.1016/j.sab.2015.12.008",
language = "English",
volume = "117",
pages = "8--15",
journal = "Spectrochimica Acta, Part B: Atomic Spectroscopy",
issn = "0584-8547",
publisher = "Elsevier",

}

TY - JOUR

T1 - Solid sampling determination of lithium and sodium additives in microsamples of yttrium oxyorthosilicate by high-resolution continuum source graphite furnace atomic absorption spectrometry

AU - Laczai, Nikoletta

AU - Kovács, L.

AU - Péter, A.

AU - Bencs, László

PY - 2016/3/1

Y1 - 2016/3/1

N2 - Solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (SS-HR-CS-GFAAS) methods were developed and studied for the fast and sensitive quantitation of Li and Na additives in microsamples of cerium-doped yttrium oxyorthosilicate (Y2SiO5:Ce) scintillator materials. The methods were optimized for solid samples by studying a set of GFAAS conditions (i.e., the sample mass, sensitivity of the analytical lines, and graphite furnace heating programs). Powdered samples in the mass range of 0.099-0.422 mg were dispensed onto graphite sample insertion boats, weighed and analyzed. Pyrolysis and atomization temperatures were optimized by the use of single-element standard solutions of Li and Na (acidified with 0.144 mol/L HNO3) at the Li I 610.353 nm and Na I 285.3013 nm analytical lines. For calibration purposes, the method of standard addition with Li and Na solutions was applied. The correlation coefficients (R values) of the calibration graphs were not worse than 0.9678. The limit of detection for oxyorthosilicate samples was 20 μg/g and 80 μg/g for Li and Na, respectively. The alkaline content of the solid samples were found to be in the range of 0.89 and 8.4 mg/g, respectively. The accuracy of the results was verified by means of analyzing certified reference samples, using methods of standard (solution) addition calibration.

AB - Solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (SS-HR-CS-GFAAS) methods were developed and studied for the fast and sensitive quantitation of Li and Na additives in microsamples of cerium-doped yttrium oxyorthosilicate (Y2SiO5:Ce) scintillator materials. The methods were optimized for solid samples by studying a set of GFAAS conditions (i.e., the sample mass, sensitivity of the analytical lines, and graphite furnace heating programs). Powdered samples in the mass range of 0.099-0.422 mg were dispensed onto graphite sample insertion boats, weighed and analyzed. Pyrolysis and atomization temperatures were optimized by the use of single-element standard solutions of Li and Na (acidified with 0.144 mol/L HNO3) at the Li I 610.353 nm and Na I 285.3013 nm analytical lines. For calibration purposes, the method of standard addition with Li and Na solutions was applied. The correlation coefficients (R values) of the calibration graphs were not worse than 0.9678. The limit of detection for oxyorthosilicate samples was 20 μg/g and 80 μg/g for Li and Na, respectively. The alkaline content of the solid samples were found to be in the range of 0.89 and 8.4 mg/g, respectively. The accuracy of the results was verified by means of analyzing certified reference samples, using methods of standard (solution) addition calibration.

KW - Advanced materials

KW - Alkaline elements

KW - Direct solid sampling analysis

KW - HR-CS-AAS

KW - Refractory matrix components

UR - http://www.scopus.com/inward/record.url?scp=84954171215&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84954171215&partnerID=8YFLogxK

U2 - 10.1016/j.sab.2015.12.008

DO - 10.1016/j.sab.2015.12.008

M3 - Article

AN - SCOPUS:84954171215

VL - 117

SP - 8

EP - 15

JO - Spectrochimica Acta, Part B: Atomic Spectroscopy

JF - Spectrochimica Acta, Part B: Atomic Spectroscopy

SN - 0584-8547

ER -