Runaway pacemaker syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker

Adam Böhm, László Hajdu, Arnold Pintér, Gábor Duray, István Préda

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

A case of a runaway pacemaker with intermittent nonoutput failure of a VVI lntermedics pacemaker is presented. This is the first report of such a phenomenon in that type of pacemaker with no apparent reason other than approaching battery end-of-life. The reported case warrants a close follow-up of patients with the same type of device.

Original languageEnglish
Pages (from-to)2143-2144
Number of pages2
JournalPACE - Pacing and Clinical Electrophysiology
Volume23
Issue number12 I
DOIs
Publication statusPublished - Jan 1 2000

Keywords

  • Nonoutput failure
  • Runaway pacemaker syndrome

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint Dive into the research topics of 'Runaway pacemaker syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker'. Together they form a unique fingerprint.