ZUR BEEINFLUSSING DER ZAHLRATENSTATISTIK DURCH DIE TOTZEIT DES STRAHLUNGSMESSGERATES

Translated title of the contribution: On the influence of radiation detector dead time upon count rate statistics

Research output: Contribution to journalArticle

Abstract

The influence of radiation detector dead time upon count rate statistics [x,s(2,x)] was investigated and it could be shown that the variance at 10,000 cps and 10 μsec dead time is 8% smaller than expected by Poisson's law.

Original languageGerman
Pages (from-to)141
Number of pages1
JournalNUC Compact
Volume11
Issue number4
Publication statusPublished - 1980

Fingerprint

Radiation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

ZUR BEEINFLUSSING DER ZAHLRATENSTATISTIK DURCH DIE TOTZEIT DES STRAHLUNGSMESSGERATES. / Csirik, J.

In: NUC Compact, Vol. 11, No. 4, 1980, p. 141.

Research output: Contribution to journalArticle

@article{6cb8a38a16ef427b97a4f66ae338cc1f,
title = "ZUR BEEINFLUSSING DER ZAHLRATENSTATISTIK DURCH DIE TOTZEIT DES STRAHLUNGSMESSGERATES",
abstract = "The influence of radiation detector dead time upon count rate statistics [x,s(2,x)] was investigated and it could be shown that the variance at 10,000 cps and 10 μsec dead time is 8{\%} smaller than expected by Poisson's law.",
author = "J. Csirik",
year = "1980",
language = "German",
volume = "11",
pages = "141",
journal = "NUC Compact",
issn = "0344-3752",
number = "4",

}

TY - JOUR

T1 - ZUR BEEINFLUSSING DER ZAHLRATENSTATISTIK DURCH DIE TOTZEIT DES STRAHLUNGSMESSGERATES

AU - Csirik, J.

PY - 1980

Y1 - 1980

N2 - The influence of radiation detector dead time upon count rate statistics [x,s(2,x)] was investigated and it could be shown that the variance at 10,000 cps and 10 μsec dead time is 8% smaller than expected by Poisson's law.

AB - The influence of radiation detector dead time upon count rate statistics [x,s(2,x)] was investigated and it could be shown that the variance at 10,000 cps and 10 μsec dead time is 8% smaller than expected by Poisson's law.

UR - http://www.scopus.com/inward/record.url?scp=0019271711&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0019271711&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:0019271711

VL - 11

SP - 141

JO - NUC Compact

JF - NUC Compact

SN - 0344-3752

IS - 4

ER -