Lowering the limits of detection of X‐ray fluorescence analysis in the electron microscope

A. Barna, I. Pozsgai, C. E. Fiori, S. A. Wight

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

A transmission‐type attachment for energy‐dispersive x‐ray fluorescence analysis was developed. The attachment was designed for analysis to be carried out in a CAMEBAX microanalyzer, where the energy‐dispersive spectrometer detector views the specimen through a hole in the objective lens and the space between the specimen and the pole piece of the objective lens is fairly small. Nevertheless, this type of attachment can be applied in other microanalyzers or scanning electron microscopes. Detection limits between 0.5 and 5 ppm on SRM 612 glass (NIST) and the applicability to the analysis of small particles (SRM 470 glass sphere 90 μm in diameter) are demonstrated.

Original languageEnglish
Pages (from-to)32-35
Number of pages4
JournalX-Ray Spectrometry
Volume23
Issue number1
DOIs
Publication statusPublished - 1994