Investigations on the photoelectron conversion efficiency of NaI(Tl) counters

M. Alflen, Zs Kajcsos, H. Spiering, P. Gütlich, R. Albrecht, R. Schulze, H. Vilhyalmsson

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

The photoelectron conversion efficiency of NaI(Tl) detectors has been measured by counting single photoelectron pulses of the detector output pulse train on a very fast storage oscilloscope. This result is compared with those obtained by the measurement of the achievable time resolution using the first photoelectron triggering technique.

Original languageEnglish
Pages (from-to)464-467
Number of pages4
JournalNuclear Inst. and Methods in Physics Research, A
Volume265
Issue number3
DOIs
Publication statusPublished - Mar 15 1988

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Dive into the research topics of 'Investigations on the photoelectron conversion efficiency of NaI(Tl) counters'. Together they form a unique fingerprint.

  • Cite this