Geometrical accuracy test obtained on the OUR rotating γ-system

Research output: Contribution to journalArticle

Abstract

In 1996, a new γ-ray brain surgical apparatus, the OUR rotating γ- system (RGS), was developed for stereotactic radiosurgery. The device contains 30 cobalt-60 sources distributed in 5 groups along 6 helical lines in a hemispherical primary collimator body, which rotates during treatment together with the built-in secondary collimator body. By rotating the primary and secondary collimator body together, 30 nonoverlapping, full 360°arcs are formed. The 30 cobalt-60 γ-radiation sources have approximately 6000 Ci total initial activity with ≥3 Gy/min initial dose rate at the isocenter. The purpose of this study was to test the geometrical accuracy of this new radiosurgical equipment. The investigations were carried out on Gafchromic Dosimetry Media, Type MD-55 films. Following a pin-prick directed toward the center point of the radiation unit, the films were exposed, and the optical density profiles were analyzed with an automatic densitometer (Molecular Dynamics Personal Densitometer, Model PD SN 50238, Innovative Optical Systems for Biological Research). By measuring the asymmetry of the position of the pin-prick in relation to the density distribution, the geometrical accuracy of the OUR RGS was determined. The calculated difference between the mechanical unit center point and the radiological center point verify the unit of ±0.1054 mm against the generally accepted ±0.5 mm.

Original languageEnglish
Pages (from-to)181-188
Number of pages8
JournalJournal of Radiosurgery
Volume2
Issue number3
DOIs
Publication statusPublished - Jan 1 1999

    Fingerprint

Keywords

  • Radiosurgery
  • Rotating γ-system
  • γ-knife

ASJC Scopus subject areas

  • Oncology
  • Clinical Neurology
  • Radiology Nuclear Medicine and imaging

Cite this