Experimental verification of an improved prefilter model

M. Simon, S. Bohátka, I. Futó, Cs Trajber

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The experimental verification of predictions of a new simulation model of quadrupole mass filters is presented. Computer calculations show that the transmission of the prefilter‐QMS system is periodical as a function of the prefilter length. The periodicity of the transmission is independent of the mass number of ions of equal velocity. Experimental proof of the simulation has been produced for low mass numbers and a more difficult procedure is required for higher masses (m/z > 100). This is realized with a closed ion source and tandem quadrupole mass spectrometer arrangement. Experimental results prove the mass‐independent behaviour predicted by the model.

Original languageEnglish
Pages (from-to)629-633
Number of pages5
JournalRapid Communications in Mass Spectrometry
Volume9
Issue number8
DOIs
Publication statusPublished - Jan 1 1995

Fingerprint

Mass spectrometers
Ion sources
Ions

ASJC Scopus subject areas

  • Analytical Chemistry
  • Spectroscopy
  • Organic Chemistry

Cite this

Experimental verification of an improved prefilter model. / Simon, M.; Bohátka, S.; Futó, I.; Trajber, Cs.

In: Rapid Communications in Mass Spectrometry, Vol. 9, No. 8, 01.01.1995, p. 629-633.

Research output: Contribution to journalArticle

Simon, M. ; Bohátka, S. ; Futó, I. ; Trajber, Cs. / Experimental verification of an improved prefilter model. In: Rapid Communications in Mass Spectrometry. 1995 ; Vol. 9, No. 8. pp. 629-633.
@article{af0c85f082fc43c181c038613a390fd8,
title = "Experimental verification of an improved prefilter model",
abstract = "The experimental verification of predictions of a new simulation model of quadrupole mass filters is presented. Computer calculations show that the transmission of the prefilter‐QMS system is periodical as a function of the prefilter length. The periodicity of the transmission is independent of the mass number of ions of equal velocity. Experimental proof of the simulation has been produced for low mass numbers and a more difficult procedure is required for higher masses (m/z > 100). This is realized with a closed ion source and tandem quadrupole mass spectrometer arrangement. Experimental results prove the mass‐independent behaviour predicted by the model.",
author = "M. Simon and S. Boh{\'a}tka and I. Fut{\'o} and Cs Trajber",
year = "1995",
month = "1",
day = "1",
doi = "10.1002/rcm.1290090802",
language = "English",
volume = "9",
pages = "629--633",
journal = "Rapid Communications in Mass Spectrometry",
issn = "0951-4198",
publisher = "John Wiley and Sons Ltd",
number = "8",

}

TY - JOUR

T1 - Experimental verification of an improved prefilter model

AU - Simon, M.

AU - Bohátka, S.

AU - Futó, I.

AU - Trajber, Cs

PY - 1995/1/1

Y1 - 1995/1/1

N2 - The experimental verification of predictions of a new simulation model of quadrupole mass filters is presented. Computer calculations show that the transmission of the prefilter‐QMS system is periodical as a function of the prefilter length. The periodicity of the transmission is independent of the mass number of ions of equal velocity. Experimental proof of the simulation has been produced for low mass numbers and a more difficult procedure is required for higher masses (m/z > 100). This is realized with a closed ion source and tandem quadrupole mass spectrometer arrangement. Experimental results prove the mass‐independent behaviour predicted by the model.

AB - The experimental verification of predictions of a new simulation model of quadrupole mass filters is presented. Computer calculations show that the transmission of the prefilter‐QMS system is periodical as a function of the prefilter length. The periodicity of the transmission is independent of the mass number of ions of equal velocity. Experimental proof of the simulation has been produced for low mass numbers and a more difficult procedure is required for higher masses (m/z > 100). This is realized with a closed ion source and tandem quadrupole mass spectrometer arrangement. Experimental results prove the mass‐independent behaviour predicted by the model.

UR - http://www.scopus.com/inward/record.url?scp=85024799348&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024799348&partnerID=8YFLogxK

U2 - 10.1002/rcm.1290090802

DO - 10.1002/rcm.1290090802

M3 - Article

AN - SCOPUS:85024799348

VL - 9

SP - 629

EP - 633

JO - Rapid Communications in Mass Spectrometry

JF - Rapid Communications in Mass Spectrometry

SN - 0951-4198

IS - 8

ER -