Digitized signal acquisition of gas-filled, multiwire, neutron detector with delay-line readout

J. Orban, L. Rosta, A. Nagy

Research output: Contribution to journalConference article

1 Citation (Scopus)

Abstract

Multiwire counters are likely to remain a reliable and cost-efficient option for a large class of instruments. Delay-line position encoding is widely used in multiwire position sensitive particle detectors. Improvement of several detector parameters is a continuous demand such as count rate or position resolution. Digitization of the analog detector signals retains all information about the detected events at the highest count rate. Neutronic measurement results on our standard detector with 200mm × 200mm active area are presented with different acquisition algorithms. The analog signals were digitized by a high speed, multichannel Acqiris digitizer with 1 ns sampling resolution.

Original languageEnglish
Article number012044
JournalJournal of Physics: Conference Series
Volume528
Issue number1
DOIs
Publication statusPublished - Jan 1 2014
EventInternational Workshop on Neutron Optics and Detectors, NOP and D 2013 - Munich, Germany
Duration: Jul 2 2013Jul 5 2013

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Fingerprint Dive into the research topics of 'Digitized signal acquisition of gas-filled, multiwire, neutron detector with delay-line readout'. Together they form a unique fingerprint.