Vízminták rádium- és radonkoncentrációjának meghatározása nyomdetektoros módszerrel

Translated title of the contribution: Determination of dissolved radon and radium content of water samples by track etch method

József Hakl, Ilona Hunyadi, István Csige, Attila Vásárhelyi, János Somlai

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

We have developed a method to determine dissolved radon and radium content in water samples using track etch type radon monitors. The device is an immersed, small volume radon monitor with CR-39 track etch detector. The monitor is sealed from the water by a thin radon permeable rubber foil. The dissolved radon and/or the radium content of water samples can be determined from two independent radon exposures by using the solutions of non steady state differential equation which describes the temporal variation of the radon content of water and by using an experimentally determined calibration coefficient.

Translated title of the contributionDetermination of dissolved radon and radium content of water samples by track etch method
Original languageHungarian
Pages (from-to)224-226
Number of pages3
JournalMagyar Kemiai Folyoirat, Kemiai Kozlemenyek
Volume103
Issue number5
Publication statusPublished - May 1 1997

ASJC Scopus subject areas

  • Chemistry(all)

Fingerprint Dive into the research topics of 'Determination of dissolved radon and radium content of water samples by track etch method'. Together they form a unique fingerprint.

  • Cite this