Calibration of CR-39-based thoron progeny device

F. Fábián, A. Csordás, A. Shahrokhi, J. Somlai, T. Kovács

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Radon isotopes and their progenies have proven significant role in respiratory tumour formation. In most cases, the radiological effect of one of the radon isotopes (thoron) and its progenies has been neglected together with its measurement technique; however, latest surveys proved that thoron's existence is expectable in flats and in workplace in Europe. Detectors based on different track detector measurement technologies have recently spread for measuring thoron progenies; however, the calibration is not yet completely elaborated. This study deals with the calibration of the track detector measurement method suitable for measuring thoron progenies using different devices with measurement techniques capable of measuring several progenies (Pylon AB5 and WLx, Sarad EQF 3220). The calibration factor values related to the thoron progeny monitors, the measurement uncertainty, reproducibility and other parameters were found using the calibration chamber. In the future, the effects of the different parameters (aerosol distribution, etc.) will be determined.

Original languageEnglish
Article numberncu072
Pages (from-to)169-172
Number of pages4
JournalRadiation Protection Dosimetry
Volume160
Issue number1-3
DOIs
Publication statusPublished - 2014

Fingerprint

radon isotopes
progeny
Calibration
Equipment and Supplies
Radon
Isotopes
Aerosols
Workplace
Uncertainty
detectors
pylons
Technology
CR 39
monitors
Neoplasms
aerosols
tumors
chambers

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Radiation
  • Public Health, Environmental and Occupational Health
  • Medicine(all)

Cite this

Calibration of CR-39-based thoron progeny device. / Fábián, F.; Csordás, A.; Shahrokhi, A.; Somlai, J.; Kovács, T.

In: Radiation Protection Dosimetry, Vol. 160, No. 1-3, ncu072, 2014, p. 169-172.

Research output: Contribution to journalArticle

Fábián, F, Csordás, A, Shahrokhi, A, Somlai, J & Kovács, T 2014, 'Calibration of CR-39-based thoron progeny device', Radiation Protection Dosimetry, vol. 160, no. 1-3, ncu072, pp. 169-172. https://doi.org/10.1093/rpd/ncu072
Fábián, F. ; Csordás, A. ; Shahrokhi, A. ; Somlai, J. ; Kovács, T. / Calibration of CR-39-based thoron progeny device. In: Radiation Protection Dosimetry. 2014 ; Vol. 160, No. 1-3. pp. 169-172.
@article{24d4c43d19294a2a94ed163a7a3b23df,
title = "Calibration of CR-39-based thoron progeny device",
abstract = "Radon isotopes and their progenies have proven significant role in respiratory tumour formation. In most cases, the radiological effect of one of the radon isotopes (thoron) and its progenies has been neglected together with its measurement technique; however, latest surveys proved that thoron's existence is expectable in flats and in workplace in Europe. Detectors based on different track detector measurement technologies have recently spread for measuring thoron progenies; however, the calibration is not yet completely elaborated. This study deals with the calibration of the track detector measurement method suitable for measuring thoron progenies using different devices with measurement techniques capable of measuring several progenies (Pylon AB5 and WLx, Sarad EQF 3220). The calibration factor values related to the thoron progeny monitors, the measurement uncertainty, reproducibility and other parameters were found using the calibration chamber. In the future, the effects of the different parameters (aerosol distribution, etc.) will be determined.",
author = "F. F{\'a}bi{\'a}n and A. Csord{\'a}s and A. Shahrokhi and J. Somlai and T. Kov{\'a}cs",
year = "2014",
doi = "10.1093/rpd/ncu072",
language = "English",
volume = "160",
pages = "169--172",
journal = "Radiation Protection Dosimetry",
issn = "0144-8420",
publisher = "Oxford University Press",
number = "1-3",

}

TY - JOUR

T1 - Calibration of CR-39-based thoron progeny device

AU - Fábián, F.

AU - Csordás, A.

AU - Shahrokhi, A.

AU - Somlai, J.

AU - Kovács, T.

PY - 2014

Y1 - 2014

N2 - Radon isotopes and their progenies have proven significant role in respiratory tumour formation. In most cases, the radiological effect of one of the radon isotopes (thoron) and its progenies has been neglected together with its measurement technique; however, latest surveys proved that thoron's existence is expectable in flats and in workplace in Europe. Detectors based on different track detector measurement technologies have recently spread for measuring thoron progenies; however, the calibration is not yet completely elaborated. This study deals with the calibration of the track detector measurement method suitable for measuring thoron progenies using different devices with measurement techniques capable of measuring several progenies (Pylon AB5 and WLx, Sarad EQF 3220). The calibration factor values related to the thoron progeny monitors, the measurement uncertainty, reproducibility and other parameters were found using the calibration chamber. In the future, the effects of the different parameters (aerosol distribution, etc.) will be determined.

AB - Radon isotopes and their progenies have proven significant role in respiratory tumour formation. In most cases, the radiological effect of one of the radon isotopes (thoron) and its progenies has been neglected together with its measurement technique; however, latest surveys proved that thoron's existence is expectable in flats and in workplace in Europe. Detectors based on different track detector measurement technologies have recently spread for measuring thoron progenies; however, the calibration is not yet completely elaborated. This study deals with the calibration of the track detector measurement method suitable for measuring thoron progenies using different devices with measurement techniques capable of measuring several progenies (Pylon AB5 and WLx, Sarad EQF 3220). The calibration factor values related to the thoron progeny monitors, the measurement uncertainty, reproducibility and other parameters were found using the calibration chamber. In the future, the effects of the different parameters (aerosol distribution, etc.) will be determined.

UR - http://www.scopus.com/inward/record.url?scp=84904883131&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84904883131&partnerID=8YFLogxK

U2 - 10.1093/rpd/ncu072

DO - 10.1093/rpd/ncu072

M3 - Article

C2 - 24723185

AN - SCOPUS:84904883131

VL - 160

SP - 169

EP - 172

JO - Radiation Protection Dosimetry

JF - Radiation Protection Dosimetry

SN - 0144-8420

IS - 1-3

M1 - ncu072