Apparent efficiency of serially coupled columns in isocratic and gradient elution modes

Szabolcs Fekete, Santiago Codesido, Serge Rudaz, Davy Guillarme, K. Horváth

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

The goal of this work was to understand the variation of apparent efficiency when serially coupling columns with identical stationary phase chemistries, but with differences in their kinetic performance. For this purpose, a mathematical treatment was developed both for isocratic and gradient modes to assess the change in plate numbers and peak widths when coupling arbitrary several columns. To validate the theory, experiments were also carried out using various mixtures of compounds, on columns packed with different particle sizes, to mimic highly efficient (new, not used) and poorly efficient columns (used one with many injections). Excellent agreement was found between measured and calculated peak widths. The average error in prediction was about 5% (which may be explained by the additional volume of the coupling tubes). In isocratic mode, the plate numbers are not additive when the coupled columns possess different efficiencies, and a limiting plate count value can be calculated depending on the efficiency and length of the individual columns. Theoretical efficiency limit can also be determined assuming one column in the row with infinite efficiency. In gradient elution mode, the columns’ order has a role (non-symmetrical system). When the last column has high enough efficiency, the gradient band compression effect may outperform the competing band broadening caused by dispersive and diffusive processes (peak sharpening). Therefore, in gradient mode, the columns should generally be sequentially placed according to their increasing efficiency.

Original languageEnglish
JournalJournal of Chromatography A
DOIs
Publication statusAccepted/In press - Jan 1 2018

Fingerprint

Particle Size
Injections
Therapeutics
Particle size
Kinetics
Experiments

Keywords

  • Apparent efficiency
  • Column coupling
  • Column length
  • Peak capacity
  • Plate number

ASJC Scopus subject areas

  • Analytical Chemistry
  • Biochemistry
  • Organic Chemistry

Cite this

Apparent efficiency of serially coupled columns in isocratic and gradient elution modes. / Fekete, Szabolcs; Codesido, Santiago; Rudaz, Serge; Guillarme, Davy; Horváth, K.

In: Journal of Chromatography A, 01.01.2018.

Research output: Contribution to journalArticle

Fekete, Szabolcs ; Codesido, Santiago ; Rudaz, Serge ; Guillarme, Davy ; Horváth, K. / Apparent efficiency of serially coupled columns in isocratic and gradient elution modes. In: Journal of Chromatography A. 2018.
@article{12469ce1bec84d19b244ff9ce5de6851,
title = "Apparent efficiency of serially coupled columns in isocratic and gradient elution modes",
abstract = "The goal of this work was to understand the variation of apparent efficiency when serially coupling columns with identical stationary phase chemistries, but with differences in their kinetic performance. For this purpose, a mathematical treatment was developed both for isocratic and gradient modes to assess the change in plate numbers and peak widths when coupling arbitrary several columns. To validate the theory, experiments were also carried out using various mixtures of compounds, on columns packed with different particle sizes, to mimic highly efficient (new, not used) and poorly efficient columns (used one with many injections). Excellent agreement was found between measured and calculated peak widths. The average error in prediction was about 5{\%} (which may be explained by the additional volume of the coupling tubes). In isocratic mode, the plate numbers are not additive when the coupled columns possess different efficiencies, and a limiting plate count value can be calculated depending on the efficiency and length of the individual columns. Theoretical efficiency limit can also be determined assuming one column in the row with infinite efficiency. In gradient elution mode, the columns’ order has a role (non-symmetrical system). When the last column has high enough efficiency, the gradient band compression effect may outperform the competing band broadening caused by dispersive and diffusive processes (peak sharpening). Therefore, in gradient mode, the columns should generally be sequentially placed according to their increasing efficiency.",
keywords = "Apparent efficiency, Column coupling, Column length, Peak capacity, Plate number",
author = "Szabolcs Fekete and Santiago Codesido and Serge Rudaz and Davy Guillarme and K. Horv{\'a}th",
year = "2018",
month = "1",
day = "1",
doi = "10.1016/j.chroma.2018.08.002",
language = "English",
journal = "Journal of Chromatography",
issn = "0021-9673",
publisher = "Elsevier",

}

TY - JOUR

T1 - Apparent efficiency of serially coupled columns in isocratic and gradient elution modes

AU - Fekete, Szabolcs

AU - Codesido, Santiago

AU - Rudaz, Serge

AU - Guillarme, Davy

AU - Horváth, K.

PY - 2018/1/1

Y1 - 2018/1/1

N2 - The goal of this work was to understand the variation of apparent efficiency when serially coupling columns with identical stationary phase chemistries, but with differences in their kinetic performance. For this purpose, a mathematical treatment was developed both for isocratic and gradient modes to assess the change in plate numbers and peak widths when coupling arbitrary several columns. To validate the theory, experiments were also carried out using various mixtures of compounds, on columns packed with different particle sizes, to mimic highly efficient (new, not used) and poorly efficient columns (used one with many injections). Excellent agreement was found between measured and calculated peak widths. The average error in prediction was about 5% (which may be explained by the additional volume of the coupling tubes). In isocratic mode, the plate numbers are not additive when the coupled columns possess different efficiencies, and a limiting plate count value can be calculated depending on the efficiency and length of the individual columns. Theoretical efficiency limit can also be determined assuming one column in the row with infinite efficiency. In gradient elution mode, the columns’ order has a role (non-symmetrical system). When the last column has high enough efficiency, the gradient band compression effect may outperform the competing band broadening caused by dispersive and diffusive processes (peak sharpening). Therefore, in gradient mode, the columns should generally be sequentially placed according to their increasing efficiency.

AB - The goal of this work was to understand the variation of apparent efficiency when serially coupling columns with identical stationary phase chemistries, but with differences in their kinetic performance. For this purpose, a mathematical treatment was developed both for isocratic and gradient modes to assess the change in plate numbers and peak widths when coupling arbitrary several columns. To validate the theory, experiments were also carried out using various mixtures of compounds, on columns packed with different particle sizes, to mimic highly efficient (new, not used) and poorly efficient columns (used one with many injections). Excellent agreement was found between measured and calculated peak widths. The average error in prediction was about 5% (which may be explained by the additional volume of the coupling tubes). In isocratic mode, the plate numbers are not additive when the coupled columns possess different efficiencies, and a limiting plate count value can be calculated depending on the efficiency and length of the individual columns. Theoretical efficiency limit can also be determined assuming one column in the row with infinite efficiency. In gradient elution mode, the columns’ order has a role (non-symmetrical system). When the last column has high enough efficiency, the gradient band compression effect may outperform the competing band broadening caused by dispersive and diffusive processes (peak sharpening). Therefore, in gradient mode, the columns should generally be sequentially placed according to their increasing efficiency.

KW - Apparent efficiency

KW - Column coupling

KW - Column length

KW - Peak capacity

KW - Plate number

UR - http://www.scopus.com/inward/record.url?scp=85052104072&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85052104072&partnerID=8YFLogxK

U2 - 10.1016/j.chroma.2018.08.002

DO - 10.1016/j.chroma.2018.08.002

M3 - Article

C2 - 30150120

AN - SCOPUS:85052104072

JO - Journal of Chromatography

JF - Journal of Chromatography

SN - 0021-9673

ER -