Accuracy studies with carbon clusters at the Penning trap mass spectrometer TRIGA-TRAP

J. Ketelaer, T. Beyer, K. Blaum, M. Block, K. Eberhardt, M. Eibach, F. Herfurth, C. Smorra, Sz Nagy

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Extensive cross-reference measurements of well-known frequency ratios using various sizes of carbon cluster ions 12Cn + (10≤n≤23) were performed to determine the effects limiting the accuracy of mass measurements at the Penning-trap facility TRIGA-TRAP. Two major contributions to the uncertainty of a mass measurement have been identified. Fluctuations of the magnetic field cause an uncertainty in the frequency ratio due to the required calibration by a reference ion of ufref)/νref = 6(2) × 10-11/min × Δt. A mass-dependent systematic shift of the frequency ratio of epsilonm(r)/r = -2.2(2) × 10-9 × (m-m ref)/u has been found as well. Finally, the nuclide 197Au was used as a cross-check since its mass is already known with an uncertainty of 0.6 keV.